·WBC 5-part differentiation,25 reportable parameters,2 histograms and 2 scattergrams
·Whole blood mode,predilute mode
·Sheath fluid+Tri-angle laser scatter+Chemical dye+Flow cytometry technology dedicated optical counting channel,10.4 inch color touch screen,with user-friendly software
·Large storage capacity:up to 50,000 sample messages,including histograms and patient information
·Testing speed:not less than 55 samples per hour
·Sample volume is only 28μL
※Characteristic
■Strong compatibility
1) Multiple sampling modes:the whole blood mode,predilute mode
2) Multiple test mode:CBC,CBC+5diff,CBC+5diff+RRBC
3) The external laser or inkjet printer
4) 4 USB ports,1LAN port supporting HL7 protocol and compatible with the hospital’s LIS system
5) Supporting quality control materials
■Low consumption and High efficiency
1) Only 3 reagents are needed
2) Sampling volume≤28μL
3) 55 samples per hour
■Humanized design
1) 10.4" inch color touch screen
2) Liquid crystal screen(LCD)
3) Resolution: 1024×768
4) 10° tilt to adapt visual observation
■Easy maintenance
1) Automatic fault handling function
2) Automatic cleaning function
■Dimension
1) 430*330*510mm
2) Weight ≤25kg
※Specification
■Principle
Coulter principle for RBC and PLT counting; HGB concentration was measured by colorimetry.
Flow cytometry was used to obtain the 5-part statistical count of WBC.
On this basis,the analyzer calculates the results of the remaining parameters.
■Parameters
25 reportable parameters:
WBC, LYM#,MON#,NEU#,EOS#,BAS#,LYM%,MON%,NEU%, EOS%,BAS%, RBC, HGB,HCT,MCV,MCH,MCHC,RDW-CV,
RDW-SD, PLT, MPV,PDW,PCT,P-LCR,P-LCC
2 histograms for RBC and PLT
2 scattergrams for Size and Diff
■Reagent
Lyse B-5LH,Lyse B-DIFF,Diluent B-DI,Probe cleanser
■Sample Volume
Whole blood mode:28μL
Predilute mode:20μL
■Print or upload LIS
If autoprint is set to on,the analyzer will automatically print the analysis report as set. If automatic communication is set to on, the analyzer automatically uploads the results of sample analysis and sample and patient information that are eligible for communication to the LIS system.
■Interface
USB,LAN
Support uni-directional LIS
Optional bar code scanner
Printer(optional)
4 USB interface and host interface:a network interface.
※Performance index
Parameters | Range of linearity | Allowable deviation range | Range of detection | Precision/% |
WBC | 1x109/L-10.0x109/L | No more than±0.5x109 /L | 3.5x109/L-9.5x109/L | ≤2.0 |
10.1x109/L-99.9x109/L | No more than±5% | |||
RBC | 0.30x1012 /L-1.00x1012 /L | No more than±0.05x1012 /L | 3.8x1012/L-5.8x1012/L | ≤1.5 |
1.01x1012 /L-7.00x1012 /L | No more than±5% | |||
HGB | 20g/L-70g/L | No more than±2g/L | 115g/L-175/L | ≤1.5 |
71g/L-200g/L | No more than±3% | |||
PLT | 20x109 /L-100x109 /L | No more than±10x109 /L | 125x109/L-350x109/L | ≤6.0 |
101x109 /L-999x109 /L | No more than±10% |
*The linear correlation coefficientr:≥0.990
■Communication
LAN port supports HL7 protocol
■Touch screen
10.4 inch color touch screen
Resolution:1024x768
■Language
Chinese,English
■Power requirement
100-240V-(±10%)
50Hz/60Hz(±1Hz)
■Operating Environmental
Temperature:10℃-35℃
Humidity:20%-80%
Air pressure:80kPa-106kPa